Open original FDA record ↗Recall record
- FDA identifier
- Z-0452-2024
- Company in FDA record
- IHB OPERATIONS B.V.
- FDA classification date
- Dec 1, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Weakened connection point on the adjustable suspension arm, potentially causing damage or a break.
Source facts
- Action
- A recall notification issued October 09, 2023 via certified mail . Follow-up will be made by service representatives within 2 weeks to arrange for correction to replace all bearing assemblies in the field with an updated design. Prior to the field correction, inspect your Freedom 42A Lightweight Spring Arm for any movement in the end bearing assembly. If any loose connections are identified, please immediately remove the Spring Arm from use and notify your local authorized representative. Address questions or concerns to your local authorized representative or Skytron at 1-800-759-8766.
- Initiated
- 2023-07-19
- Root cause
- Device Design
- Distribution
- US: AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MB, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, VT, WA, WI, WV.
- Recall number
- Z-0452-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204067
- FDA classified
- 2023-12-01
- Recall event number
- 93351