Open original FDA record ↗Recall record
- FDA identifier
- Z-0452-2026
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Nov 7, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in temperature during clinical use.
Source facts
- Action
- On October 8, 2025 URGENT: MEDICAL DEVICE ADVISORY NOTICE letters were sent to customers. Actions to be taken: Ensure you read the content of the notice. Ensure that all personnel are aware and knowledgeable on the content of the notice. You may continue to use the device according to the instruction for use, which cautions users to ensure that a back-up transducer and probe are sterilized and available prior to beginning a procedure. If you have further distributed this product, identify your customers, and forward them this notification. Olympus requests that you acknowledge receipt of this letter with the provided reply form.
- Initiated
- 2025-10-08
- Root cause
- Device Design
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, India, Taiwan, Germany, Australia, Singapore.
- Recall number
- Z-0452-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216278
- FDA classified
- 2025-11-07
- Recall event number
- 97665