There is a remote possibility for the plastic latch to deform over the lifetime of use. When this happens there is the possibility for the seat to become dislodged from the base. And when not using the positioning strap, there is a risk that the occupant could slide out of the seat and sustain injury.
Source facts
Action
Urgent Medical Device Field correction letters were sent to Sunrise Suppliers/Dealers and Zippie Customers on November 2013 by first class mail. Foreign notification was supplied to suppliers and regulatory affairs director in Canada.
Initiated
2013-11-11
Root cause
Nonconforming Material/Component
Terminated
2014-03-18
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, Australia, Hong Kong, and China.
Recall number
Z-0453-2014
Recall status
Terminated
CDRH record ID
124071
FDA classified
2013-12-05
Recall event number
66896