The firm identified a nonconformance in Lot 04882 involving the endcap being manufactured to a longer length than indicated in product specifications, introducing the risk of the +6 mm endcap not seating flush with the nail while inserting during a procedure.
Source facts
Action
The firm indicated that they will send recall notification to sales representatives via email or certified mail. Customers will be instructed to remove all +6 mm endcaps from stock and to return to the recalling firm with a return label that will be provided at a later date. Customers with questions are to contact the recalling firm at (678)235-3314.
Initiated
2021-11-17
Root cause
Device Design
Distribution
Domestic: CA, FL, GA, IL, IN, MA, MD, TX, and WA.
Recall number
Z-0453-2022
Recall status
Open, Classified
CDRH record ID
190571
FDA record created
2022-01-04
Recall event number
89117