Open original FDA record ↗Recall record
- FDA identifier
- Z-0454-05
- Company in FDA record
- Boston Scientific Scimed
- FDA classification date
- Jan 28, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Runway guide catheter do not meet internal product specifications. Specifically, the affected catheters have exposed or protruding braid wires in the inner lumen of the catheter tip. This condition could have the potential to damage a POBA or Stent Delivery System balloon
Source facts
- Action
- On 10/29/04 the Boston Scientific Sales force was notified via voice mail of the recall. Sales personnel were told to contact their accounts/customers by the end of the business day on 11/02/04. Accounts/customers were to receive written notice of the recall via Fed Ex on Monday 11/01/04. Sales Personnel are to ensure that accounts/customers got notice of the recall, understand instructions and are taking appropriate steps. Catheters are to be returned to Boston Scientific Corporation, Quincy, MA.
- Initiated
- 2004-10-29
- Root cause
- Other
- Terminated
- 2006-11-10
- Distribution
- Throughout United States and into 15 other countries.
- Recall number
- Z-0454-05
- Recall status
- Terminated
- CDRH record ID
- 35814
- FDA classified
- 2005-01-28
- Recall event number
- 30405