Open original FDA record ↗Recall record
- FDA identifier
- Z-0454-2009
- Company in FDA record
- Beckman Coulter Inc
- FDA classification date
- Dec 16, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect results: under certain circumstances, when transfer volume is not within the set range, the Vidiera NsP software does not report sample exclusion and does not flag the excluded sample in the "Run Results" report. The software error associated with the excluded sample appears as "Liquid was not successfully transferred to well#..........the sample will be excluded from further proce
Source facts
- Action
- The customer notifications were initiated by Beckman Coulter on September 11, 2007 when all affected accounts were contacted via telephone, followed by a Product Corrective Action (PCA) was mailed via US Postal Service to all affected customers. The PCA informed the customers of the issue and instructed them to: (1) Review the "run Summary" event log for your samples. (2) Review the attached instructions describing the steps for identifying the samples affected by this error. (3) Complete and return the enclosed response form within 10 days so that we may maintain our records. The letter informed the customers that Beckman Coulter would provide a software update within 60 days. The letter also asked their customers to share the PCA with their staff and provide a copy of the PCA to any customers that they may have forwarded the affected products to. For questions regarding the PCA, please contact firm at 714-961-3634.
- Initiated
- 2007-08-30
- Root cause
- Software design
- Terminated
- 2009-01-13
- Distribution
- Nationwide distribution including states of PA, MI, UT, WA, IL, and CA.
- Recall number
- Z-0454-2009
- Recall status
- Terminated
- CDRH record ID
- 71828
- FDA classified
- 2008-12-16
- Recall event number
- 48660