Open original FDA record ↗Recall record
- FDA identifier
- Z-0455-2008
- Company in FDA record
- Abbott Laboratories
- FDA classification date
- Mar 13, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect Results: Hemoglobin results in donor samples assayed on the CELL-DYN 1800 System vary more than expected across the range of claimed operating temperatures of 18 degrees Celcius to 30 degrees Celcius.
Source facts
- Action
- A Product Correction letter was sent to all customers with a reply form on October 27, 2006. Product information letters are being packed with the instrument's accessory kit and with calibrators and quality control products used with the CELL-DYN 1800 System. Correction letters with reply forms have been resent to nonresponding customers.
- Initiated
- 2006-10-27
- Root cause
- Software design
- Terminated
- 2011-01-06
- Distribution
- Worldwide Distribution, USA and countries of Argentina, Bolivia, Brazil, Canada, Chile, Columbia, Costa Rica, Dominican Republic, El Salvador, Germany, Guatemala, Honduras, Hong Kong, China, Japan, Mexico, Paraguay, Puerto Rico, Singapore, South Korea and Venezuela.
- Recall number
- Z-0455-2008
- Recall status
- Terminated
- CDRH record ID
- 53826
- FDA classified
- 2008-03-13
- Recall event number
- 38465