Incorrect treatment plan or record: A software design deficiency has the potential for corrupted treatment plan records or treatment records in cases where multiple users editing a record simultaneously can result in a corrupted (but accessible) treatment plan record.
Source facts
Action
On May 13, 2010, the firm issued letters to all its direct consignees, informing them of the affected products and providing instructions on the recall.
***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.***
Initiated
2010-05-13
Root cause
Software design
Terminated
2012-02-24
Distribution
Nationwide and Worldwide
Recall number
Z-0455-2011
Recall status
Terminated
CDRH record ID
92423
FDA classified
2010-11-24
Recall event number
56012