Z-0455-2014: Synthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Paralle

Recall record Z-0455-2014 for Synthes USA HQ, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0455-2014
Company in FDA record
Synthes USA HQ, Inc.
FDA classification date
Dec 5, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A potential non conformance was identified with the Zero-P VA plate where the screw could back out if the blocking mechanism is not visually confirmed to properly cover the head of the screw.

Source facts

Action
Synthes sent an email message dated March 23, 2012, to the sales representative who received the affected product. The email identified the product, the problem, and the action to be taken. The letter requested the sales rep examine inventory, stop using the product, and quarantine the affected product to ensure that it will not be used. Synthes instructed the customer not to remove any product from their account or return any product without further instruction. For questions they were instructed to call 610-719-5364. For questions regarding this recall call 610-719-5000.
Initiated
2013-09-30
Root cause
Process design
Terminated
2015-04-16
Distribution
Nationwide Distribution - only AZ .
Recall number
Z-0455-2014
Recall status
Terminated
CDRH record ID
122562
FDA classified
2013-12-05
Recall event number
66538

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗