Open original FDA record ↗Recall record
- FDA identifier
- Z-0455-2016
- Company in FDA record
- CareFusion 303, Inc.
- FDA classification date
- Dec 21, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
CareFusion is recalling the Alaris EtCO2 module model 8300 because it may have been tested during service using a higher than specified concentration of CO2. The effect to the EtCO2 module tested with a higher concentration of CO2 could lead to an incorrect measurement of CO2.
Source facts
- Action
- CareFusion sent an Urgent: Medical Device Recall Notification Important letter dated November 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The Alaris EtCO2 Module ( model 8300) is being recalled because it may have been tested during service using a higher than specified concentration of CO2 Customers with questions regarding clinical inquiries, product complaints or clinical troubleshooting are instructed to contact Customer Advocacy at (888) 812-3266 or email at customerfeedback@carefusion.com. Customers with technical questions regarding the Alaris are instructed to contact Technical Support at (888) 812-3229 or by email at DL-US-INF-TechSupport@carefusion.com.
- Initiated
- 2015-11-12
- Root cause
- Other
- Terminated
- 2017-01-23
- Distribution
- US Distribution.
- Recall number
- Z-0455-2016
- Recall status
- Terminated
- CDRH record ID
- 141712
- FDA classified
- 2015-12-21
- Recall event number
- 72632