Z-0455-2022: QIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223

Recall record Z-0455-2022 for Qiagen Sciences LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0455-2022
Company in FDA record
Qiagen Sciences LLC
FDA record creation date
Jan 5, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an affected cartridge is used, false-negative, false-positive, or both false-negative and false-positive results could occur.

Source facts

Action
On November 8, 2021, the firm began notifying customers. Customers received an Urgent Medical Device Correction Notice via email. Customers were instructed to do the following: - Please check the cartridge SNs received from LOT 210209 against the list of affected cartridges shown in Table 1 (page 1). If you have received one of the affected cartridges, please contact QIAGEN Technical Service for a free-of-charge replacement. - If you have received one of the 16 affected cartridges and have not used it, please dispose of it immediately in accordance with your national and local safety and environmental regulations. - If you already used one of the affected cartridges from this LOT, please identify the results obtained with the affected SNs listed on Table 1 (page 1). For results obtained with an affected SN, review the results as follows: o Review all results of the respective targets mentioned on page 1 to exclude erroneous diagnosis and treatment, except in those cases where alternative confirmation was obtained. o Review any result for which tests were performed on clinical and/or epidemiological suspicion of the respective targets. - If you have remaining stocks from LOT 210209, cat. no 691223 with cartridge SNs not found in the list, you may still use these cartridges, as this issue is limited to the 16 cartridges listed on Table 1 (page 1). - Review this notice with your laboratory/medical director. - IMPORTANT: Forward this information to all individuals and departments within your organization using the above listed kits. If you are not the end user, please forward this notice to the product end user. - Complete the Acknowledgement of Receipt attached to the letter by November 22, 2021. If you have any questions or concerns, please contact your local QIAGEN Technical Service Department through any of the following: Telephone: 800 362 7737 Email: TechService-NA@qiagen.com
Initiated
2021-11-08
Root cause
Employee error
Terminated
2023-03-29
Distribution
Domestic distribution to AL, DC, KY, MD, ME, NJ, OK< PA, and TX. No foreign distribution.
Recall number
Z-0455-2022
Recall status
Terminated
CDRH record ID
190588
FDA record created
2022-01-05
Recall event number
89130

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗