Open original FDA record ↗Recall record
- FDA identifier
- Z-0456-2021
- Company in FDA record
- Merit Medical Systems, Inc.
- FDA record creation date
- Nov 16, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The flocked tip of the 100mm swab may break off in the nasophyaryngeal cavity during sample collection. This could result in soft tissue injury of the nasophyaryngeal cavity or a delay in treatment.
Source facts
- Action
- The firm sent an recall notification on September 16, 2020 to all affected consignees. In addition to informing consignees about the correction, the firm requested consignees take the following action: 1. Please refer to the attached instructions and/or linked instructional video when collecting samples with nasopharyngeal swabs. 2. Ensure that applicable personnel within your organization are made aware of this notice. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 4. If you have any questions or concerns regarding this communication or the products addressed, please contact your Merit Sales Representative or Merit Customer Service at (801) 208-4381.
- Initiated
- 2020-09-16
- Root cause
- Device Design
- Terminated
- 2021-06-01
- Distribution
- Worldwide distribution. US Nationwide, Australia, Canada, Colombia, Ghana, Ireland, Jamaica, Kenya, Panama, Qatar, Trinidad and Tobago.
- Recall number
- Z-0456-2021
- Recall status
- Terminated
- CDRH record ID
- 183906
- FDA record created
- 2020-11-16
- Recall event number
- 86522