Open original FDA record ↗Recall record
- FDA identifier
- Z-0457-2009
- Company in FDA record
- Radiometer America Inc
- FDA classification date
- Dec 3, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The firm found that for this particular lot that the barcode, used for entering the characteristics of CAL1 Solution into the analyzer, does not reflect the actual values of the solution. As a consequence, the calibration curves for these parameters and thereby the measured results for patient samples and quality controls will be biased.
Source facts
- Action
- On, 9/2/2008 the firm sent out recall notification letters requesting their customers discard any Calibration 1 Solution (CAL1) S1720 (part number 944-024), lot WA-04, complete the attached return form and fax it to 800-736-0601. The firm will send out replacements for all affected product that the firm shows was shipped to their location. Further questions could be directed to Technical Support at 800-736-0600, option 2.
- Initiated
- 2008-09-05
- Root cause
- Process control
- Terminated
- 2012-04-25
- Distribution
- Worldwide Distribution --- USA including states of AZ, CA, FL, IA, ID, IL, IN, KY, LA, MD, NC, ND, NE, NY, OH, OK, PA, SD, TX, VA, WA, and WI, and country of Canada.
- Recall number
- Z-0457-2009
- Recall status
- Terminated
- CDRH record ID
- 73497
- FDA classified
- 2008-12-03
- Recall event number
- 49468