Open original FDA record ↗Recall record
- FDA identifier
- Z-0457-2016
- Company in FDA record
- BC Group International Inc
- FDA classification date
- Dec 22, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Device is not functioning as intended: Two wires running to the ECG distribution circuit board are reversed resulting in inverted waveforms.
Source facts
- Action
- The recalling firm issued return authorization numbers to the two customers who submitted complaints regarding the malfunctioning device. The customers sent back the units and the recalling firm replaced their devices with new units. On 10/29/2015, the recalling firm contacted their third customer by phone and informed them that the three devices they had in their possession were miswired which would cause the inverted ECG waveform. The recalling firm issued a return authorization number to their customer for the three devices. The recalling firm told their customer they would repair the devices and return them.
- Initiated
- 2015-10-28
- Root cause
- Employee error
- Terminated
- 2017-02-15
- Distribution
- US Nationwide Distribution in the states of South Carolina, Florida, and Canada.
- Recall number
- Z-0457-2016
- Recall status
- Terminated
- CDRH record ID
- 141675
- FDA classified
- 2015-12-22
- Recall event number
- 72619