Open original FDA record ↗Recall record
- FDA identifier
- Z-0457-2021
- Company in FDA record
- Stryker GmbH
- FDA record creation date
- Nov 16, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The device has the potential to release nickel above the acceptable margin of safety in pediatric patients who weigh less than 20 kg (44.09 lbs) when 2 or more implants are used.
Source facts
- Action
- All affected customers were mailed an Urgent Medical Device Correction (UMDC) letter on 09/25/2020 that requested them to take the following actions: 1. Please inform users of this Urgent Medical Device Correction (UMDC) and forward this notice to all individuals who need to be made aware of the recall. 2. Under 21 CFR 803, manufacturers are also required to report any serious injuries where a product has contributed or may have contributed to the event. Please keep Stryker informed of any adverse events associated with this product by emailing soprodexpreports@stryker.com. 3. A response is required, even though you may not have any physical inventory on site anymore. 4. Hospitals/Branches/Agencies: Complete and sign the enclosed Urgent Medical Device Correction Business Reply Form and fax a copy to 1-888-240-4294 or email to StrykerOrtho5568@Stericycle.com. 5. Branches/Agencies Only: Please forward this Urgent Medical Device Correction to the individuals or organizations who have consigned product, if applicable. Please assist us in meeting our regulatory obligation by faxing back the attached Urgent Medical Device Correction Business Reply Form within 5 days. We regret any inconvenience this action may cause you and if you have any questions or concerns after reviewing the letter, please contact Customer Service at (201) 831-5000. For questions pertaining to the recall, email SO_M_PRODUCT_FIELD_ACTION_RESPONSE@stryker.com
- Initiated
- 2020-09-25
- Root cause
- Under Investigation by firm
- Distribution
- The devices were distributed nationwide throughout the US, and to countries OUS. Easy Clip devices: As of 09/04/2020, 251,566 EasyClip devices have been manufactured. 121,517 of those devices were distributed to the US, and 102,529 were distributed in countries OUS. 27,520 devices were left in the firm's inventory. Easy Clip Xpress devices: As of 09/04/2020, 37,371 EasyClip Xpress devices have been manufactured. 9,561 of those devices were distributed to the US, leaving 27,810 in the firm's inventory. This device is only distributed in the United States.
- Recall number
- Z-0457-2021
- Recall status
- Open, Classified
- CDRH record ID
- 184140
- FDA record created
- 2020-11-16
- Recall event number
- 86597