Z-0457-2021: Osteosynthesis Compression Staple EasyClip

Recall record Z-0457-2021 for Stryker GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0457-2021
Company in FDA record
Stryker GmbH
FDA record creation date
Nov 16, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The device has the potential to release nickel above the acceptable margin of safety in pediatric patients who weigh less than 20 kg (44.09 lbs) when 2 or more implants are used.

Source facts

Action
All affected customers were mailed an Urgent Medical Device Correction (UMDC) letter on 09/25/2020 that requested them to take the following actions: 1. Please inform users of this Urgent Medical Device Correction (UMDC) and forward this notice to all individuals who need to be made aware of the recall. 2. Under 21 CFR 803, manufacturers are also required to report any serious injuries where a product has contributed or may have contributed to the event. Please keep Stryker informed of any adverse events associated with this product by emailing soprodexpreports@stryker.com. 3. A response is required, even though you may not have any physical inventory on site anymore. 4. Hospitals/Branches/Agencies: Complete and sign the enclosed Urgent Medical Device Correction Business Reply Form and fax a copy to 1-888-240-4294 or email to StrykerOrtho5568@Stericycle.com. 5. Branches/Agencies Only: Please forward this Urgent Medical Device Correction to the individuals or organizations who have consigned product, if applicable. Please assist us in meeting our regulatory obligation by faxing back the attached Urgent Medical Device Correction Business Reply Form within 5 days. We regret any inconvenience this action may cause you and if you have any questions or concerns after reviewing the letter, please contact Customer Service at (201) 831-5000. For questions pertaining to the recall, email SO_M_PRODUCT_FIELD_ACTION_RESPONSE@stryker.com
Initiated
2020-09-25
Root cause
Under Investigation by firm
Distribution
The devices were distributed nationwide throughout the US, and to countries OUS. Easy Clip devices: As of 09/04/2020, 251,566 EasyClip devices have been manufactured. 121,517 of those devices were distributed to the US, and 102,529 were distributed in countries OUS. 27,520 devices were left in the firm's inventory. Easy Clip Xpress devices: As of 09/04/2020, 37,371 EasyClip Xpress devices have been manufactured. 9,561 of those devices were distributed to the US, leaving 27,810 in the firm's inventory. This device is only distributed in the United States.
Recall number
Z-0457-2021
Recall status
Open, Classified
CDRH record ID
184140
FDA record created
2020-11-16
Recall event number
86597

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗