Z-0457-2022: Simplexa Direct Amplification Disc Kit, Product Kit Code MOL1455

Recall record Z-0457-2022 for DiaSorin Molecular LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0457-2022
Company in FDA record
DiaSorin Molecular LLC
FDA record creation date
Jan 5, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A subset of affected discs are at risk of leakage when it is used more than one time with Simplexa COVID-19 Direct and/or Simplexa Flu A/B & RSV Direct Gen II. Leakage from a well that had a positive sample may cause further cross-contamination in areas of the lab if not properly cleaned.

Source facts

Action
On 11/19/2021, the firm notified affected customers via email of "URGENT: MEDICAL DEVICE CORRECTION NOTIFICATION." The letter indicated, "Actions to be taken by the Customer/User/Distributor: - Any remaining discs in inventory that are listed in the table below should only be used one time and not reused. - Any sample with an error code should be retested. - If leakage is suspected, the spillage should be removed and the instrument surface should be decontaminated. Ensure the practice of Good Laboratory Procedures including regular cleaning. - Notify others within your network that may have received these kits. - Complete and return the Acknowledgement and Receipt Form (page 3) of this notification letter. - If your firm has further distributed the devices, please share a copy of this correction notification and response form with those customers and provide a copy of the completed response form back to us. If you have any questions or if you would like replacement product for the Direct Amplification Discs you still have in stock, please contact DiaSorin Molecular Technical Service, Monday Friday, 7:00am to 5:00pm (Pacific Time) at (800) 838-4548, option 3, or by email at technicalinfo.molecular@DiaSorin.com."
Initiated
2021-11-19
Root cause
Under Investigation by firm
Distribution
US: AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV; and Worldwide: Kuwait, Australia, Israel, Italy, Canada, Chile, Puerto Rico, United Arab Emirates.
Recall number
Z-0457-2022
Recall status
Open, Classified
CDRH record ID
190485
FDA record created
2022-01-05
Recall event number
89098

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗