A malfunction was detected with one of the components that, under rare circumstances, may result in the overheating of the POD. This overheating, while unlikely to occur, has the potential to cause a burn if the POD is in direct contact with the patient.
Source facts
Action
An Urgent: Medical Device Recall letter was emailed to customers on 10/23/2009. The letter described the issue and hazard and asked that the use of the device be discontinued and to immediately contact Nonin for return and repair/replacement instructions. Questions or concerns should be directed to John Dalpee at 763-577-3166.
Initiated
2009-10-23
Root cause
Device Design
Terminated
2012-03-29
Distribution
Worldwide Distribution -- USA, including states of LA, MN, NC, and NY and countries of Canada, Germany, Italy, Netherlands, and United Kingdom.
Recall number
Z-0458-2010
Recall status
Terminated
CDRH record ID
86537
FDA classified
2009-11-25
Recall event number
53801