Vibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support
Source facts
Action
Siemens issued on 10/19/18 an Urgent Medial Devoce Correction notification to via AX033/18/S advising users of the problem, health risk, steps to be taken.
Siemens has initiated screw connection exchange of the ceiling bearing via Update Instruction AXOSS/17 /S.
Initiated
2018-10-19
Root cause
Device Design
Terminated
2020-04-27
Distribution
US Nationwide Distribution
Recall number
Z-0458-2019
Recall status
Terminated
CDRH record ID
168866
FDA record created
2018-11-14
Recall event number
81514