False Negative Results: Air bubbles in the lens or the area surrounding the lens of the LightCycler 100 uL capillaries may cause false negative results. The affected capillaries showed up to a 40% reduction in signal
Source facts
Action
Consignees were notified by Urgent Medical Device Recall letter dated 8/30/07, to discontinue use of the 100uL capillaries, plus all of those received at installation, and to disposed of them per local guidelines.
Initiated
2007-08-30
Root cause
Process design
Terminated
2008-09-08
Distribution
Nationwide
Recall number
Z-0459-2008
Recall status
Terminated
CDRH record ID
54521
FDA classified
2007-12-12
Recall event number
38897