Open original FDA record ↗Recall record
- FDA identifier
- Z-0460-2026
- Company in FDA record
- Sonesta Medical AB
- FDA classification date
- Nov 7, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Video fluoroscopy table may contain inadequate weld on backrest bracket resulting in backrest dropping.
Source facts
- Action
- On September 8, 2025, Sonesta Medical AB issued a recall notification to affected consignees via E-Mail. Sonesta asked consignees to take the following actions: 1. Please review your table and confirm that the serial number on the product labeling matches with the number listed in the notification. 2. Please discontinue use of the table. It is recommended that you place a copy of this notice on the table to provide staff awareness of the issue. 3. Sonesta will contact you to schedule and perform a table inspection as soon as possible. 4. If the table inspection results are acceptable, Sonesta will provide you with evidence to continue to use the table. 5. If the table inspection results are unacceptable, Sonesta and Laborie will work with you to replace your table If you have any questions, contact the CEO via email: Pelin.Sari@sonestamedical.se or phone +468-502 572 80.
- Initiated
- 2025-08-22
- Root cause
- Process control
- Distribution
- US Distribution to NH only.
- Recall number
- Z-0460-2026
- Recall status
- Open, Classified
- CDRH record ID
- 215957
- FDA classified
- 2025-11-07
- Recall event number
- 97568