Open original FDA record ↗Recall record
- FDA identifier
- Z-0461-2012
- Company in FDA record
- Life Technologies Corporation
- FDA classification date
- Jan 11, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
In vitro diagnostic reagent may be contaminated with a fungal contaminant.
Source facts
- Action
- Life Technologies sent an URGENT DEVICE RECALL letter dated November 14, 2011 by mail to all affected customers. The letter identified the product, description of problem, and actions to be taken by the user. Customers were instructed to examine inventory immediately, discontinue use, quarantine the material and return remaining inventory. Customers were advised to complete the Customer Response Sheet included for recovering inventory data and fax back to (+1) 716-774-6727 Technical Support or e-mail to techsupport@lifetech.com. Customers who had distributed the affected kits outside of their facility were instructed to immediately notify their customers of the recall. For questions call Technical Support, at (+1) 800-955-6288 opt. 2 then opt. 3.
- Initiated
- 2011-11-14
- Root cause
- Material/Component Contamination
- Terminated
- 2012-02-10
- Distribution
- Worldwide Distribution - USA (nationwide) in the states of CA and NV and the countries of: Canada, Venezuela, and the UK.
- Recall number
- Z-0461-2012
- Recall status
- Terminated
- CDRH record ID
- 105719
- FDA classified
- 2012-01-11
- Recall event number
- 60542