- Class II
Reason stated in the source
The firm is issuing this voluntary removal following receipt of reports of parastomal mesh failure identified several years following parastomal hernia repair using the modified Sugarbaker repair technique. In these reports, product failure led to hernia recurrence requiring additional surgical treatment. Symptoms of hernia recurrence may include discomfort, localized pain-free or painful bulging, and possible changes in the overlying skin. Medtronic has received, worldwide, a total of ten reports of mesh failure following use of the product in the last five years. Patients who have received a Parietex" composite parastomal mesh for the treatment of a parastomal hernia should continue to receive ongoing monitoring by their healthcare providers for the recurrence of a parastomal hernia.
Source facts
- Action
- On October 5, 2018, the firm notified its customers of the product issue via Urgent Medical Device Voluntary Removal. Required Actions: 1. Please quarantine and discontinue use of the affected item codes. 2. Complete the Product Return Form even if you do not have inventory. If no inventory, fax to 203-492-7719 or email FCAMITG@covidien.com. 3. If purchased directly from Medtronic, fax the return form to 800-895-6140 or email feedback.customerservice@covidien.com. Ship affected product with RGA# provided by customer service to: Medtronic Attn: Field Returns Department 195 McDermott Road North Haven, CT 06473 USA 4. If purchased from a distributor, contact distributor directly to arrange for return of product. Fax completed form to 203-492-7719 or email FCAMITG@covidien.com. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
- Initiated
- 2018-10-05
- Root cause
- Unknown/Undetermined by firm
- Terminated
- 2020-05-07
- Distribution
- Worldwide distribution--US Nationwide
- Recall number
- Z-0461-2019
- Recall status
- Terminated
- CDRH record ID
- 168668
- FDA record created
- 2018-11-14
- Recall event number
- 81417