Skip to content

Public FDA record

Z-0461-2021: cobas infinity central lab/cobas infinity core license-The cobas infinity IT solutions application is int

Recall record Z-0461-2021 for Roche Diagnostics Operations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0461-2021
Company in FDA record
Roche Diagnostics Operations, Inc.
FDA record creation date
Nov 17, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential Incorrect Validation of Results Due to an Erroneous QC Status When Using Status Expiration Control Rule

Source facts

Action
Roche issued Urgent Medical Device Correction TP-01100 is on 10/9/20 via UPS Ground (receipt signature required). Letter states reason for recall, health risk and action to take: Until cobas infinity Service Patches are available and implemented at your site, follow the containment measures outlined in the Applying Containment Measures section of this UMDC. " If you run the QC module in the cobas infinity application, contact the Roche Support Network Customer Support Center at 1-800-526-2272, and a representative will run a script to help you determine if your facility might be impacted. " Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. " Complete the attached faxback form (TP-01111) and fax or email it according to the instructions on the form. " File this Urgent Medical Device Correction (UMDC) for future reference. The cobas infinity Service Patches 3.01.12 (to correct affected versions 3.01.xx) and 3.02.05 (to correct affected versions 3.02.xx) are planned to be available in the fourth quarter of 2020. Once the Services Patches are available, a Roche Diagnostics representative will contact you to coordinate the corresponding installation.
Initiated
2020-10-09
Root cause
Software design
Terminated
2023-06-09
Distribution
Nationwide
Recall number
Z-0461-2021
Recall status
Terminated
CDRH record ID
184190
FDA record created
2020-11-17
Recall event number
86609

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗