Recall record Z-0462-2021 for LivaNova USA Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0462-2021
Company in FDA record
LivaNova USA Inc
FDA record creation date
Nov 17, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to an extra digit being inadvertently added to the serial number of a subset of implantable pulse generators, they become not compatible with therapy programmers with software version 11.0.4.

Source facts

Action
On 09/28/2020, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter to customers via certified 1st class return receipt informing them of an incompatibility warning message will appear with the use of the VNS Therapy Programmers with software version 11.0.4 upon interrogation of certain generators (specifically serial numbers e 1,000,000). This software incompatibility issue is due to the use of a now discontinued 7-digit serial number, thereby resulting in an error in the serial number and causing a communication disruption. For certain generators (SN e 1,000,000) using software version 11.0.4, a warning message will appear. Communication, programming, and diagnostic testing with this subset of affected Model devices will not be possible with software version 11.0.4. This software incompatibility issue does not affect the functionality of the generator itself. The device itself performs as intended when communication, programming and diagnostic testing is done with a compatible software version. Customers are instructed: For patients implanted with a device containing a 7-digit serial number (SN e 1,000,000): 1. Ensure patient has a Therapy Programmer with a software version that is compatible with affected Model 106 devices (SN e 1,000,000) in accordance with the Physicians Manual (including Model 250 version 11.0.3/11.0.5, all current software versions of Model 3000). 2. If you only have a Programmer with software version 11.0.4, contact the Recalling Firm's sales representative or Customer Quality to receive a Programmer that is compatible for all of your patients devices. Note: To check the software version of the Programmer, do the following: 1. Power on the Programmer 2. Review the software version provided on the Main Menu In addition, customer are asked to: 1. Ensure that the customer notification is communicated to all personnel with their organization who needs to be aware of this issue. 2. Fill out and Customer
Initiated
2020-09-28
Root cause
Process control
Terminated
2022-11-02
Distribution
U.S.: AK, AL, AZ, CA, CO, DC, FL, HI, ID, IL, IN, KS, KY, LA, ME, MI, MN, MO, MT, NC, NE, NV, OK, OR, PR, TX, UT, VA, and WA O.U.S.: Argentina, Finland and Peru
Recall number
Z-0462-2021
Recall status
Terminated
CDRH record ID
184052
FDA record created
2020-11-17
Recall event number
86570

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗