Skip to content

Public FDA record

Z-0463-2022: Alinity m System, Part No. 08N53-002

Recall record Z-0463-2022 for Abbott Molecular, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0463-2022
Company in FDA record
Abbott Molecular, Inc.
FDA record creation date
Jan 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is an issue with the installation of updated camera firmware on the system.

Source facts

Action
On December 06, 2021 Abbott Molecular issued a Product Correction letter to all customers who received an Alinity m instrument impacted by this issue (List Number 08N53 002). Customers are being informed of a previously completed assessment conducted by Abbott FSEs. Customers were initially informed at the time of the assessment, but proper reconciliation of the communication was not performed. 100% of customers are requested to return to Abbott a Customer Reply Form acknowledging receipt and understanding, and/or a request for assistance. Abbott personnel will follow up with customers by e mailing and or follow up visits/phone calls.
Initiated
2021-12-06
Root cause
Software design
Terminated
2024-09-30
Distribution
Distribution in the United States in CA, DC, GA, KY, IL, MD, MI, MN, MS, NH, NJ, NY, RI, SD, TX, WI OUS distribution to Australia, Cambodia, Canada, Columbia, France, Germany, Israel, Italy, Malaysia, Mozambique, Portugal, South Africa, Spain, Switzerland, Taiwan, United Kingdom, and Vietnam
Recall number
Z-0463-2022
Recall status
Terminated
CDRH record ID
190681
FDA record created
2022-01-06
Recall event number
89169

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗