Open original FDA record ↗Recall record
- FDA identifier
- Z-0464-03
- Company in FDA record
- Meridian Bioscience Inc
- FDA classification date
- Jan 23, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Due to defective packaging, the firm cannot assure product stability through the labeled expiration date.
Source facts
- Action
- The firm mailed notices to their affected customers on 12/26/2002. The consignees were instructed to visually and physically inspect the seals on each microwell pouch for gaps. If properly sealed, the use of the microwells may be continued. Test wells in pouches with defective seals are to be quarantined and destroyed. The customers are asked to report back to the firm. All test microwell plates in defective pouches will be replaced by the firm.
- Initiated
- 2002-12-26
- Root cause
- Other
- Terminated
- 2004-03-29
- Distribution
- The kits were distributed to end-users and distributors located nationwide; and to foreign distributors located in China, Guatemala, New Zealand, Hong Kong, Mexico, Great Britain, Belgium, Italy, Brazil, Australia, Canada, Thailand, Honduras, and Japan.
- Recall number
- Z-0464-03
- Recall status
- Terminated
- CDRH record ID
- 25559
- FDA classified
- 2003-01-23
- Recall event number
- 25295