Z-0464-2022: Instructions for Use for all model numbers of the Medtronic Abre Venous Self-Expanding Stent System.

Recall record Z-0464-2022 for Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0464-2022
Company in FDA record
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA record creation date
Jan 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There have been reports of stent migration.

Source facts

Action
The firm issued letters dated November 2021 on 11/19/2021 via 2-day UPS delivery to the U.S. consignees. The letter informs the consignee of the upcoming updates to the Instruction for Use (IFU) for the Abre venous self-expanding stent system and explains the updates will provide new information to help mitigate the risk of possible stent migration. The Customer Instructions in the letter requests the enclosed upcoming updates to the IFU be reviewed, the letter is shared with all those who need to be aware within their organization, patients should continue to be monitored per their practice's normal follow-up procedures, and the enclosed Customer Confirmation Form be completed and emailed to the firm. The updated electronic IFU will be available before the end of the calendar year. In Outside the United States (OUS) geographies, communication of the Medical Device Correction letter began 11/24/2021 via regionally approved methods. Consignees will be notified of the upcoming IFU change, as well as provided the content of the IFU change. Consignees will be asked to complete and return a confirmation certificate confirming they have received the notification. The IFU for OUS consignees will be available in the summer of 2022.
Initiated
2021-11-19
Root cause
Other
Terminated
2024-08-16
Distribution
Distribution was nationwide. There was also military/government distribution. Foreign distribution was made to Argentina, Australia, Austria, Bahrain, Belgium, Botswana, Bulgaria, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Jordan, Kenya, Lithuania, Malaysia, Republic of Moldova, Netherlands, New Zealand, Nicaragua, Norway, Oman, Peru, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, and Viet Nam,
Recall number
Z-0464-2022
Recall status
Terminated
CDRH record ID
190621
FDA record created
2022-01-06
Recall event number
89140

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗