Open original FDA record ↗Recall record
- FDA identifier
- Z-0465-2013
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Dec 3, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Firm has become aware of an unintended behavior when using syngo.plaza and issued an Update Instruction. The Update Instruction SY050/12/S provides a software update and addresses the following issues: - Patient Rename messages for the same patient are being continually processed for several days. This fills up the 'pcvdbsrv' log and potentially affects performance of the system. - Data could
Source facts
- Action
- The firm, Siemens, sent "Update Instruction SY050/12/S" on June 20, 2012 to its consignees/customers. The Update Instructions described the product, problem and actions to be taken. It also provides a software update. If you have any questions call 610-219-4834 or email: anastasia.mason@siemens.com.
- Initiated
- 2012-06-20
- Root cause
- Software design
- Terminated
- 2014-01-14
- Distribution
- Nationwide distribution: USA including states of: CA, FL, GA, MA, MN, MO, NJ, NY, OH, and WI.
- Recall number
- Z-0465-2013
- Recall status
- Terminated
- CDRH record ID
- 114137
- FDA classified
- 2012-12-03
- Recall event number
- 63521