Skip to content

Public FDA record

Z-0465-2022: Disposable Subdermal Needle Electrode, Corkscrew, Single Use, Sterile; Model Numbers TE/S50715-001-C, TE/

Recall record Z-0465-2022 for Technomed Europe. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0465-2022
Company in FDA record
Technomed Europe
FDA record creation date
Jan 7, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a high risk of the blue hub detaching from the corkscrew needle, resulting the the needle remaining in the patient's scalp.

Source facts

Action
The firm distributed an Urgent Medical Device Recall notice dated 12/07/2021 to customers in the United States via email on 12/08/2021. In this notice the firm identifies affected product as being recalled due to the high risk of the needle electrode and lead wire coming loose from the hub, resulting in potential medical intervention to remove the inserted needle electrode, including removal using forceps or incision. The recalling firm requested that customers notify any recipients of recalled product down the distribution chain by forwarding a copy of the notice, that the customer be aware that the firm will schedule a telephone call to initiate the replacement of affected product, and that only distributors return the provided Acknowledgement Form following the scheduled phone call to quality@technomed.nl. Customers are asked to ascertain the amount of affected product on hand to be returned with their lot numbers for the scheduled call. Customers with any questions are to contact the recalling firm at quality@technomed.nl or by phone at +31 43 608 48 48.
Initiated
2021-12-08
Root cause
Device Design
Terminated
2024-05-31
Distribution
US Nationwide distribution in the states of California, Florida, and Ohio.
Recall number
Z-0465-2022
Recall status
Terminated
CDRH record ID
190582
FDA record created
2022-01-07
Recall event number
89126

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗