The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.
Source facts
Action
Philips and Neusoft Medical Systems Co., Ltd. sent an Urgent Device Correction NeuViz Dual, NeuViz 16 letter, dated Oct. 18, 2012, to all affected US customers and distributors.
The letter described the product, problem, and actions to be taken by the customers. Customers are advised to:
1. Have a mechanic check the patient table top screws fixing. Detailed
instructions are attached as Annex with the letter.
2. If it is found that table top screws are improperly assembled, stop using
the product and contact local Service personnel for repair.
Philips and Neusoft Medical Systems Co., Ltd is in the process of upgrading devices with aforementioned kits and will contact customers to arrange for an update to affected systems.
If you have any questions about this corrective action program, please contact the Service Support Department, Neusoft Medical Systems Co. Ltd. at nms-service@neusoft.com
For questions regarding this recall call 425-487-7665.
Initiated
2012-10-18
Root cause
Device Design
Terminated
2013-07-08
Distribution
Nationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico.
Recall number
Z-0466-2013
Recall status
Terminated
CDRH record ID
114197
FDA classified
2012-12-04
Recall event number
63553