Z-0466-2013: NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System

Recall record Z-0466-2013 for Philips And Neusoft Medical Systems Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0466-2013
Company in FDA record
Philips And Neusoft Medical Systems Co., Ltd.
FDA classification date
Dec 4, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.

Source facts

Action
Philips and Neusoft Medical Systems Co., Ltd. sent an Urgent Device Correction NeuViz Dual, NeuViz 16 letter, dated Oct. 18, 2012, to all affected US customers and distributors. The letter described the product, problem, and actions to be taken by the customers. Customers are advised to: 1. Have a mechanic check the patient table top screws fixing. Detailed instructions are attached as Annex with the letter. 2. If it is found that table top screws are improperly assembled, stop using the product and contact local Service personnel for repair. Philips and Neusoft Medical Systems Co., Ltd is in the process of upgrading devices with aforementioned kits and will contact customers to arrange for an update to affected systems. If you have any questions about this corrective action program, please contact the Service Support Department, Neusoft Medical Systems Co. Ltd. at nms-service@neusoft.com For questions regarding this recall call 425-487-7665.
Initiated
2012-10-18
Root cause
Device Design
Terminated
2013-07-08
Distribution
Nationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico.
Recall number
Z-0466-2013
Recall status
Terminated
CDRH record ID
114197
FDA classified
2012-12-04
Recall event number
63553

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗