Z-0466-2016: Zimmer Oscillating Flat - TS ST Hub; 75mm x 19mm x 1.00/1.27mm; Synvasive Part No. 11-4770: Part No. 19

Recall record Z-0466-2016 for Synvasive Technology Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0466-2016
Company in FDA record
Synvasive Technology Inc
FDA record creation date
Dec 22, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Two lots of oscillating saw blades incorrectly identify the blade cutting width as 25mm instead of the correct specification of 19mm.

Source facts

Action
Zimmer sent an Urgent Medical Device Removal letter on December 10, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed: I. Review this notification. 2. You will receive copies of notification sent directly to hospital risk managers and surgeons in your territory. As necessary, review and facilitate understanding of this notification by those entities. 3. Further identify hospital risk managers and/or surgeons in your territory who should also receive notification of this update, and supply the information to the entities you have identified. 4. If after reviewing this notification you have further questions or concerns, please call 330-364-0989 between the hours of 8 a.m. and 5 p.m. EST, Monday through Friday. Letters to Rick Managers are similar and the actions requested by them are: 1. Reviewing this notification and forwarding to all people within your organization that may utilize this product. 2. Identifying and quarantining the affected product to prevent further distribution or use. 3. Completing the attached Response Form and returning it via e-mail to CorporateQua1ity.PostMarket@zimmerbiomet.com. Returning affected product to: Zimmer Surgical Attn: QA!RA Dept. - Recall Please include a copy of the Response Form with the shipment. 4. Upon receipt of affected product, Zimmer will credit your account for unused Zimmer Oscillating Flat Blade - TS ST Hub, 75mm x 19mm, xl.OO/l.27mm, lots 43054 and 45104 only. Please return a copy of the completed response form along with your returned product to ensure proper credit. Important: Please distribute this notification to all personnel within your organization who need to be aware. If you have further distributed affected product(s), please provide the customer 's information on the Business Response Form to Zimmer. Letters also included a FAQ sheet regarding the labeling error.
Initiated
2015-12-08
Root cause
Labeling False and Misleading
Terminated
2016-11-01
Distribution
US Distribution to one location each in CT, MA and CA.
Recall number
Z-0466-2016
Recall status
Terminated
CDRH record ID
142288
FDA record created
2015-12-22
Recall event number
72811

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗