- Z-0467-2016
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Dec 22, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Possible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the helium gas to be blown directly into the magnet room or other areas. This could lead to displacement of oxygen, cause cold burns, and/or pose a risk to users, patient or others.
Source facts
- Action
- Siemens sent an Customer Advisory letter dated December 7, 2015, certified mail to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter notifies customers of the potential injuries, which may result of improperly installed quench lines. The letter indicates it is the system's owner who is responsible for installing the quench line per the requirements. If Siemens installed the quench lines, the consignee is directed to contact Siemens, who will inspect the quench lines during next maintenance visit. Siemens will monitor service reports for affected sites. For further questions please call ( 610) 219-6300.
- Initiated
- 2015-12-07
- Root cause
- Other
- Terminated
- 2017-06-21
- Distribution
- Nationwide including : DC, PR, Virgin Islands and Bermuda
- Recall number
- Z-0467-2016
- Recall status
- Terminated
- CDRH record ID
- 142368
- FDA classified
- 2015-12-22
- Recall event number
- 72861