Z-0467-2025: Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQ

Recall record Z-0467-2025 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0467-2025
Company in FDA record
Bard Peripheral Vascular Inc
FDA classification date
Nov 15, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle detachment and misplacement, which could lead to tissue damage and bleeding.

Source facts

Action
On 9/30/24, removal notices were sent to customers who were asked to do the following: 1) Discontinue use of product listed in the letter, and immediately dispose of all affected product remaining in your possession in accordance with your local facility s process. 2) Share this notification with all users within your facility network of the product to ensure they are also aware of this field action. If the affected products have been distributed further, please share this notice to ensure that these entities are contacted and informed of the recall. 3) Complete and return the response form via email to BDRC27@BD.COM - If you purchased this product from a distributor, contact your distributor for further instructions and resolution. - Affected lots distributed from 05 April 2023 through 23 Aug 2024. If further assistance is required contact the firm: - North American Regional Complaint Center: 1-844-823-5433, Mon - Fri 8:00am and 5:00pm CT, productcomplaints@bd.com - Recall Related Questions: BDRC27@BD.COM On 1/15/2025, additional removal notices were mailed or emailed to distributors informing them of additional affected devices.
Initiated
2024-09-30
Root cause
Process change control
Distribution
Worldwide - US Nationwide including in the states of WA, LA, MN, PA, UT, SC, OK, KY, AZ, IL, TX, NJ, NY, FL, TN, IN, OH, NC, CO, VA, GA, MD, CA, AL, WI, MO, IA, CT, NE, MI, MS, NV, KS, OR, SD, ID, NH, RI, WY, AR, MA, DE, ND, NM, AK, PR and the countries of Portugal, Spain, South Africa, Greece, United Arab Emirates, United Kingdom, Sweden, Finland, Belgium, Korea, Australia, Canada.
Recall number
Z-0467-2025
Recall status
Open, Classified
CDRH record ID
210375
FDA classified
2024-11-15
Recall event number
95491

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗