Skip to content

Public FDA record

Z-0468-2011: CSS-059-40-4 Modified 4.0mm Modified Depth Gauge, Lot 0985011

Recall record Z-0468-2011 for Orthohelix Surgical Designs Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0468-2011
Company in FDA record
Orthohelix Surgical Designs Inc
FDA classification date
Nov 26, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The print for modified CSS depth gauges has inaccurate definition of the etched screw length. The error results in a length reading that is longer (5 mm in the case of the 7.0 and 4mm in the cases of the 5.5 and 4.0) than the actual length of the buried guide wire segment and even longer (10 mm in the case of the 7.0 and 8mm in the cases of the 5.5 and 4.0) than the design intention readings. This

Source facts

Action
OrthoHelix issued an Important Recall Notice letter dated October 15, 2010 to their customers, identifying the affected products. Customers were requested to immediately return the afected devices. OrthoHelix can be contacted at 330 247-1444 concerning this recall.
Initiated
2010-10-15
Root cause
Device Design
Terminated
2012-09-06
Distribution
Nationwide distribution: to the states of GA and MO.
Recall number
Z-0468-2011
Recall status
Terminated
CDRH record ID
95385
FDA classified
2010-11-26
Recall event number
57095

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗