Open original FDA record ↗Recall record
- FDA identifier
- Z-0469-2014
- Company in FDA record
- Omnyx Llc
- FDA classification date
- Dec 9, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is an issue associated with Digital Pathology System (DPS) Software, in which the annotation tools within the DPS version 1.1 Image Viewer will display incorrect measurements and area calculations for a region of interest on a scanned slide, under certain conditions.
Source facts
- Action
- Omnyx sent a Urgent Field Safety Notice letter dated October 1, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Please discontinue use of the measurement and area calculation features of the annotation tools (line, rectangle/square, ellipse/circle, and freeform) in the DPS Image Viewer at this time. If any of these tools have been used for measurements or area calculations on clinical cases, verify measurements and calculations on glass slides. Please confirm receipt of these instructions by completing the attached form and scanning/emailing it to FieldAction@ omnyx.com or faxing it to 001-412-894-2137 Product Correction: An updated version of the DPS SW will be released and installed on your system as soon as it becomes available. Please note that the appropriate regulatory authorities have been informed of this field action. We apologize for any inconvenience that this may have caused you and are working diligently to address this issue. If you have any additional questions or concerns, please contact you GE Healthcare account representative. Further questions please call (412) 891-2100.
- Initiated
- 2013-10-01
- Root cause
- Software design
- Terminated
- 2014-04-16
- Distribution
- International Distribution to Canada and Switzerland only.
- Recall number
- Z-0469-2014
- Recall status
- Terminated
- CDRH record ID
- 122675
- FDA classified
- 2013-12-09
- Recall event number
- 66596