Skip to content

Public FDA record

Z-0469-2014: Digital Pathology System (DPS) Software version 1.1 The Omnyx" Integrated Digital Pathology system, whic

Recall record Z-0469-2014 for Omnyx Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0469-2014
Company in FDA record
Omnyx Llc
FDA classification date
Dec 9, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is an issue associated with Digital Pathology System (DPS) Software, in which the annotation tools within the DPS version 1.1 Image Viewer will display incorrect measurements and area calculations for a region of interest on a scanned slide, under certain conditions.

Source facts

Action
Omnyx sent a Urgent Field Safety Notice letter dated October 1, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Please discontinue use of the measurement and area calculation features of the annotation tools (line, rectangle/square, ellipse/circle, and freeform) in the DPS Image Viewer at this time. If any of these tools have been used for measurements or area calculations on clinical cases, verify measurements and calculations on glass slides. Please confirm receipt of these instructions by completing the attached form and scanning/emailing it to FieldAction@ omnyx.com or faxing it to 001-412-894-2137 Product Correction: An updated version of the DPS SW will be released and installed on your system as soon as it becomes available. Please note that the appropriate regulatory authorities have been informed of this field action. We apologize for any inconvenience that this may have caused you and are working diligently to address this issue. If you have any additional questions or concerns, please contact you GE Healthcare account representative. Further questions please call (412) 891-2100.
Initiated
2013-10-01
Root cause
Software design
Terminated
2014-04-16
Distribution
International Distribution to Canada and Switzerland only.
Recall number
Z-0469-2014
Recall status
Terminated
CDRH record ID
122675
FDA classified
2013-12-09
Recall event number
66596

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗