Skip to content

Public FDA record

Z-0469-2015: LUNA Chest, Model No. LU4006

Recall record Z-0469-2015 for Novadaq Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0469-2015
Company in FDA record
Novadaq Technologies, Inc.
FDA classification date
Dec 1, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Non sterile-drape LUNA drape was inadvertently mislabeled with an additional contents sterile label.

Source facts

Action
Novadaq initiated their recall by telephone and followed by an email and a Medical Device Recall Notice dated November 11, 2014, to all affected customers. Distributors were instructed to notify their customers. The firm is removing the recalled products and replacing them with correctly labeled product. Customers were instructed to confirm receipt of the Recall Notice and their understanding of the actions to be taken by completing the attached confirmation form and faxing it to Novadaq Quality Department at 1-800-886-2419. For questions regarding this recall call 604-432-9861.
Initiated
2014-11-04
Root cause
Error in labeling
Terminated
2015-11-06
Distribution
Nationwide Distribution to IA, LA, GA, MD, ID, FL, TN, OH, TX, NC, MA, KY, KS, WI, IN, SC, OR, CT, CA, WA, PA, OK, and PR.
Recall number
Z-0469-2015
Recall status
Terminated
CDRH record ID
131610
FDA classified
2014-12-01
Recall event number
69839

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗