Skip to content

Public FDA record

Z-0469-2019: AIA-900 Analyzer, Product code 022930, 022930R

Recall record Z-0469-2019 for Tosoh Bioscience Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0469-2019
Company in FDA record
Tosoh Bioscience Inc
FDA record creation date
Nov 14, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Tosoh Bioscience has become aware of a potential issue with the pickup arm and pickup arm motor on the AIA-900 Analyzer. If the pickup arm or pickup arm motor malfunctions, the analyzer will cease to function and result in an inability to test patient samples. This may cause a delay in testing and reporting patient results. Tosoh has received fifty-one (51) complaints related to this issue with no serious injuries reported.

Source facts

Action
On October 5, 2018, the firm notified its affected customers via an Urgent Medical Device Recall letter. The letter informed customers of the potential product issue with all active AIA-900 analyzers. The firm's service procedures have been updated to proactively troubleshoot and correct the pickup arm and pickup arm motor during installations and scheduled preventive maintenance visits. At the next scheduled preventive maintenance visit a Tosoh field service engineer will verify the pickup arm motor current and adjust as needed. Adherence to preventive maintenance schedules provides effective mitigation against potential failure of these components. Immediate Actions to be taken by the Customer/User " Thoroughly review the content of this letter. " Continue to use the instrument. A Tosoh representative will contact you to schedule your next preventive maintenance visit. o If a failure of the pickup arm is observed or the analyzer stops functioning, contact Tosoh Technical Support for an expedited service visit. " Complete and return the attached Acknowledgement Form by any of the listed methods. " File this notice with your laboratory records and forward this information to others in your laboratory. If you have obtained unexpected test results, delays in test result reporting, or received any complaints of illness or adverse events associated with the use of the analyzer, please contact Tosoh Technical Support 24 hours a day, seven days a week at (800) 248-6764. Should you have any questions regarding this medical device recall, please feel free to contact Bernadette O'Connell at (800) 248-6764 or by email at bernadette.oconnell@tosoh.com. She will be available to answer any questions Monday - Friday, from 9:00 AM to 5:00 PM (PST).
Initiated
2018-10-05
Root cause
Device Design
Terminated
2020-11-10
Distribution
Distributed nationwide. Foreign distribution to Canada and Latin America.
Recall number
Z-0469-2019
Recall status
Terminated
CDRH record ID
167729
FDA record created
2018-11-14
Recall event number
81187

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗