Z-0469-2021: Artis zee floor, Artis Q ceiling, Artis Q.zen ceiling. Image-Intensified Fluoroscopic X-Ray System. Mater

Recall record Z-0469-2021 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0469-2021
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Nov 18, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some electronic components within the generator were equipped with an incorrect protection diode. As a result, protection against overvoltage may be inadequate, and the system may not function properly in the event an overvoltage situation occurs.

Source facts

Action
On October 21, 2020, the firm notified customers via Urgent Medical Device Correction letter. The letter informed customers of the hardware error, and that customers could continue to use their system but needed to follow their standard emergency procedures they had in place in the event of system failures. It is recommended to consider reinforcing these emergency procedures until the update has been completed. Siemens advises that an alternative system is available to continue treatment in case of occurrence. Siemens is implementing a corrective action to all affected users via Update Instruction AX018/20/S. The Siemens Service organization will exchange the affected printed circuit boards of the generator. This modification will bring the voltage back to the center of the tolerance range. Siemens' service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact the service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented.
Initiated
2020-10-21
Root cause
Employee error
Terminated
2021-05-05
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AR, CA, IA, IL and the countries of China, Germany, India, Indonesia, Italy, Japan, Philippines, Saudi Arabia, South Korea, Taiwan, Turkey, and United Kingdom.
Recall number
Z-0469-2021
Recall status
Terminated
CDRH record ID
184269
FDA record created
2020-11-18
Recall event number
86660

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗