Skip to content

Public FDA record

Z-0469-2022: Regard, T AND A - Rogers, sterilized convenience kits

Recall record Z-0469-2022 for ROi CPS LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0469-2022
Company in FDA record
ROi CPS LLC
FDA record creation date
Jan 10, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Kits lacks rubber latex warning label.

Source facts

Action
A customer notification email dated 11/30/2021 was sent with recall instructions as follows: 1. Check: Please review your inventory to determine if the manufacturer item numbers and lot numbers listed in the table are in your inventory. 2. Quarantine: If the affected kits are identified, please quarantine the product and do not further distribute. 3. Label Affected kits: Use the label template attached to this notice to label affected kits in your inventory prior to shipping to customers. The label includes the statement "This kit contains a URETHRAL CATHETER RED RUBBER 12FR that contains natural rubber latex which may cause allergic reactions." The label should be placed in a location on the outer wrapper of the kits that is visible to end users. 4. Reply: Please complete the attached reply form indicating any inventory on hand at your facility. If you do not have inventory, please return the form indicating "0" inventory on hand. 5. Notify: Please immediately notify any customers to who, you have distributed or forwarded kits affected by this medical device correction notice. You may include a copy of this notice and the attached label template with your customer notification. 6. End users: Notices sent to end users should include the label template and instruction to label affected kits in their inventory. The label should be placed in a location on the outer wrapper of the kits that is visible to users of the kits.
Initiated
2021-11-30
Root cause
Error in labeling
Terminated
2023-08-07
Distribution
US Nationwide distribution in the state of MO.
Recall number
Z-0469-2022
Recall status
Terminated
CDRH record ID
190644
FDA record created
2022-01-10
Recall event number
89149

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗