Offset Bushings associated with the lots identified were mismarked: the rotational reference numbers on the Tibial Offset Bushing were marked counterclockwise instead of clockwise; the Femoral Offset Bushing did not include rotational markings.
Source facts
Action
Stryker Orthopaedics sent an "URGENT PRODUCT RECALL" letter dated July 27, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Product Recall Acknowledgement Form was attached for customers to complete and return via fax to 201-831-6069. Contact the firm at 201-972-2100 for assistance with this recall.
Initiated
2012-07-27
Root cause
Error in labeling
Terminated
2015-09-30
Distribution
Worldwide Distribution.
Recall number
Z-0470-2013
Recall status
Terminated
CDRH record ID
114071
FDA classified
2012-12-04
Recall event number
63442