Reason stated in the source
This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the software defect is triggered when the user switches between these trajectories during a surgical procedure.
Source facts
- Action
- Synaptive Medical sent an Urgent Medical Device Recall letter dated October 9, 2018 via registered priority mail. The letter identified the affected product, problem and actions to be taken until a software patch is released. Customers were requested to complete and return the enclosed Acknowledgement and Receipt form to Synaptive Medical Inc. For questions call 1-844-462-7246.
- Initiated
- 2018-10-09
- Root cause
- Software design
- Terminated
- 2021-08-12
- Distribution
- Worldwide - US Nationwide Distribution and the country of Canada
- Recall number
- Z-0470-2019