Open original FDA record ↗Recall record
- FDA identifier
- Z-0471-2015
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA classification date
- Nov 26, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire Heart displayed as analysis results of the CFA and in a Left-Ventricular Volume Curve generated based on some of those parameters.
Source facts
- Action
- Toshiba sent an Urgent Medical Device Correction letter dated November 5, 2014, to all affected customers. Customers were asked to refrain from using the function parameters for the entire heart and the left-ventricular volume curve until corrective actions have been taken. Customers were instructed to share this information with all users and reviewing radiologist as well as clinical engineering or Biomedical group at facility. Please complete and return the attached form and fax it to the toll free number at the top of the form 877-349-3054. This form can also be sent via email to raffairs@tams.com. Revised software to prevent occurrence of this problem will be installed on the system. When the new software becomes available the Toshiba service representative will contact for an appointment to install it on the system. Customers with questions were instructed to call 800-421-1968 or contact their local Toshiba Representative at 800-521-1968.
- Initiated
- 2014-11-05
- Root cause
- Software design
- Terminated
- 2015-10-09
- Distribution
- US Nationwide Distribution to CA, MA, MD
- Recall number
- Z-0471-2015
- Recall status
- Terminated
- CDRH record ID
- 131461
- FDA classified
- 2014-11-26
- Recall event number
- 69748