Open original FDA record ↗Recall record
- FDA identifier
- Z-0473-2013
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Dec 4, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditions can lead to insufficient reagent pipetting. This may affect patient sample and controls recovery for the following tests: Alkaline Phosphatase (ALP2), Alanine Aminotransferase (ALT), Bicarbonate (CO2-L), Creatinine (C
Source facts
- Action
- On 11/01/2012 an "Urgent Medical Device Correction" notification was sent via UPS Ground (receipt required) to consignees. The device issue was described and recommended actions were provided. A Roche Field Application Specialist will install the new e-barcode during a scheduled visit. Customers also will have an option to install the new e-barcodes themselves, following the instructions provided in the notification. A Software update will be released when available. Customers may contact Technical Support at 1-800-428-2336 for more information.
- Initiated
- 2012-11-01
- Root cause
- Software design
- Terminated
- 2014-01-08
- Distribution
- Nationwide Distribution, including the states of AL, AZ, CA, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, NJ, NY, OH, VA, WA, and WI.
- Recall number
- Z-0473-2013
- Recall status
- Terminated
- CDRH record ID
- 114234
- FDA classified
- 2012-12-04
- Recall event number
- 63577