Skip to content

Public FDA record

Z-0473-2014: Synthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedge os

Recall record Z-0473-2014 for Synthes USA HQ, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0473-2014
Company in FDA record
Synthes USA HQ, Inc.
FDA classification date
Dec 10, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A labeling correction was initiated related to the Surgical Technique Guide J5606-C "TomoFix Osteotomy System" due to the potential for a surgical delay if the surgeon inserts the superior screws at a lesser angle (4 or 6 degrees) and is not aware that those angles are not present.

Source facts

Action
Synthes sent an Urgent Field Safety Notification/Medical Device Labeling Correction letter dated October 9, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to exchange the old surgical technique guide version J5606-C with the new version J5606-D provided with this notification. Review the revised surgical technique guide (Pages 25 & 26). Forward this Field Safety Notification to anyone in their facility that need to be informed. If the surgical technique guide has been forwarded to another facility, contact that facility. Maintain awareness of this notice until all surgical technique guides have been exchanged. Maintain a copy of this notice. Customers with questions were instructed to contact their Synthes Sales Consultant. For questions regarding this recall call 610-719-5000.
Initiated
2013-10-09
Root cause
Labeling False and Misleading
Terminated
2015-12-18
Distribution
Worldwide Distribution - USA (nationwide) Canada and Mexico
Recall number
Z-0473-2014
Recall status
Terminated
CDRH record ID
122651
FDA classified
2013-12-10
Recall event number
66582

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗