The items in this lot are missing the ion implantation feature. The surface of the head may be softer and more prone to scratching, which in tum could potentially cause higher wear of the poly bearing. Increased wear could potentially lead to a shorter implant life.
Source facts
Action
Biomet, Inc. sent an URGENT MEDICAL DEVICE RECALL NOTICE dated November 1, 2012 to all affected consignees. The letter identified the affected product, problem and actions to be taken. The notification letter instructed the surgeon to montior the patient as their is a potential for increased wear which have an unknown effect on implant life. For questions call 574-372-1570.
Initiated
2012-11-01
Root cause
Process control
Terminated
2013-08-29
Distribution
USA Nationwide Distribution in the state of WI.
Recall number
Z-0474-2013
Recall status
Terminated
CDRH record ID
114592
FDA classified
2012-12-05
Recall event number
63708