Z-0475-06: COULTER GEN.S Series Hematology Analyzer, PN 6605632, 6605632R; GENS: 6605381, 6605381R, 6605360, 6605360

Recall record Z-0475-06 for Beckman Coulter Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0475-06
Company in FDA record
Beckman Coulter Inc
FDA classification date
Feb 4, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

There is a risk of sample misidentification when processing samples in the Manual aspiration mode.

Source facts

Action
A Product Corrective Action letter will be mailed the week of Dec 19, 2005 to all COUL TER@ LH700 Series System & COUL TER@ GEN.STM System customers instructing them of a rare risk of sample misidentification. (Letter will be sent by US mail)
Initiated
2005-12-15
Root cause
Other
Terminated
2012-01-26
Distribution
Nationwide and Canada
Recall number
Z-0475-06
Recall status
Terminated
CDRH record ID
43740
FDA classified
2006-02-04
Recall event number
34380

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗