An oversight in the implementation of the design change record resulted in outdated components being shipped to customers. The component did not pass a half-cycle sterilization.
Source facts
Action
The distributors were verbally notified, and replacement Rev F Hex Drivers sent to the locations of the kits. The kits had been moved in preparation for upcoming surgeries, but the Drivers were replaced prior to any additional surgery taking place. No affected product remains in the field. For further questions, please call (508) 520-3003.
Initiated
2014-06-09
Root cause
Process change control
Terminated
2016-02-05
Distribution
US Distribution to the states of : PA, WV and CA.
Recall number
Z-0475-2016
Recall status
Terminated
CDRH record ID
141991
FDA classified
2015-12-23
Recall event number
72695