Open original FDA record ↗Recall record
- FDA identifier
- Z-0475-2021
- Company in FDA record
- Conformis, Inc.
- FDA record creation date
- Nov 20, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Incorrect hip components were provided in kits.
Source facts
- Action
- The firm received complaints regarding the mix-up of serial numbers on September 11, 2020, and September 21, 2020, from a sales representative who reported that the incorrect iView bearing serial number 0469255 was provided in the kit for serial number 0469225, and that the incorrect iJigs bearing serial number 0469225 were provided in the kit for serial number 0469255. One of the kits was used, while the other was returned to the firm. The firm reported that the co-mingle only involved these two kits. Upon notification of the issue, the firm implemented a planned deviation to verify that there were no other occurrences of the issue. All components were found to be associated with the correct serial number. If you have any questions, contact Director, Regulatory Affairs at 781-345-9001 or email: kara.johnson@conformis.com.
- Initiated
- 2020-09-21
- Root cause
- Process control
- Terminated
- 2023-06-12
- Distribution
- US Distribution in CA.
- Recall number
- Z-0475-2021
- Recall status
- Terminated
- CDRH record ID
- 184198
- FDA record created
- 2020-11-20
- Recall event number
- 86615