Z-0476-2023: Leica MICROSYSTEMS Provido, Part Numbers 10448950 (Premium), 10448976 (Standard). Surgical microscope

Recall record Z-0476-2023 for Leica Microsystems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0476-2023
Company in FDA record
Leica Microsystems, Inc.
FDA classification date
Dec 9, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.

Source facts

Action
The firm issued an URGENT FIELD SAFETY NOTICE to its consignees by email on 11/21/2022. The notice explained the issue and stated that a Leica Microsystems representative would be contacting each facility to arrange an update to the the affected devices and advised the following: Consignees can continue to use the system pending the field update; Ensure that all users of the affected microscope are aware of the Safety notes as described in section 3 of the respective user manual; Like any other instrument in the operating theater, the system may fail and Leica Microsystem (Schweiz) AG recommends that a backup system be kept available during the operation; Check before the operation : Connect the power cable. + Switch on the microscope. UPDATE 09/12/2023: Recall was expanded, and customers were sent the notification on about 09/11/2023. Each consignee is requested to acknowledge receipt of the Field Safety Notice letter outlining the potential manufacturing defect, perform an initial inspection, indicate the results of the initial inspection, and provide contact information for further Leica Microsystems Representative verification of conformity/defect.
Initiated
2022-11-21
Root cause
Process control
Distribution
US Nationwide
Recall number
Z-0476-2023
Recall status
Open, Classified
CDRH record ID
197249
FDA classified
2022-12-09
Recall event number
91200

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗