Z-0477-2019: Binding Site Human IgM Kit (for SPAPLUS), REF: NK012.S The device consists of reaction buffer, sheep ant

Recall record Z-0477-2019 for The Binding Site Group, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0477-2019
Company in FDA record
The Binding Site Group, Ltd.
FDA classification date
Nov 17, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

After running the calibration curve, the recovery of the IgM kit control results decrease demonstrating a reduction in accuracy within the same day; control results may be out of range one day after running the assay calibration curve. Samples have been shown to be similarly affected.

Source facts

Action
On 07/25/2018, the Urgent Field Safety Notice and Field Correction E-back Form were emailed to customers. Customers were advised to cease use of the affected product and to dispose of the kits locally. This notice states that a witness signature of kit disposal is required on the E-back form. This notice needs to be passed on to all those who need to be aware within your organisation or to any organisation where the potentially affected devices have been transferred. Customers with additional questions are encouraged to contact their local firm representative, or customer service via phone at +44(0) 1214569696, or email at Technical.support@bindingsite.co.uk. The E-back Form requests that each customer and witness sign and date confirming they have both received and have followed the advice issued in the Notification. It also asks for the number of product in stock and number of kits disposed.
Initiated
2018-07-25
Root cause
Under Investigation by firm
Terminated
2021-04-27
Distribution
U.S.: IL, WV, NE, SD, NY, IN, WA, MD, NJ, PA, NC, AR, FL, VA
Recall number
Z-0477-2019
Recall status
Terminated
CDRH record ID
168646
FDA classified
2018-11-17
Recall event number
81409

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗